| Beamline Website |
http://sls.web.psi.ch/view.php/beamlines/px/index.html |
| Owner/Operator |
Paul Scherrer Institut |
| Operational (general use) |
June 2006 |
| %Time available (general use) |
75 |
| Beamline Schedule |
http://sls.web.psi.ch/view.php/organization/status/scheduled/index.html?BEAMLINE=PX&POS=4013
|
| Experiments |
Monochromatic, MAD |
| Source |
In-vacuum Undulator |
| Ring Current(mA) |
300 - 400 mA |
| Average Beam Lifetime |
infinite |
| Spot Size |
Micro-diffractometer: 25 x 5 m**2; High-Resolution D: 85 x 10 m**2 |
| Optics |
sagittally horizontal focussing crystal; meridionally vertical focussing mirror |
| Flux |
> 1E12 at all wavelength |
| Flux Details |
highest flux at 1.5 A |
| Wavelength Min (Å) |
0.71 |
| Wavelength Max (Å) |
2.15 |
| Monochromator |
Fixed-exit LN2 cooled Double Crystal Monochromator |
| Crystal type |
Si(111) |
| Mirrors |
Rh coated meridionally focussing mirror |
| Mirror Cutoff |
18 keV |
| Special Capabilities |
Focussing on the detector; low resolution data collection |
| Other Instrumentation |
SAXS tube for data down to 1000 ; robot under preparation |
| Goniometer |
single phi axis, 4DX arc (18 mm pin length) |
| Pin Length |
any |
| Restrictions |
no |
| xtal-detector Distance |
80 - 1200 |
| Detector |
mar225 CCD, PILATUS 6M |
| Detector Details |
The MAR 225 CCD is on the mu diffractometer. The PILATUS 6M is on the high resolution diffractometer.
|
| 2theta range |
0 - 45 |
| Cryo Capability |
N2, He at prior request |
| Lab Facilities |
small beamline lab, incubator, 16 - 25 room, 4 room |
| Biohazard |
no |
| Software |
xds, mosflm, ccp4, shelx, sharp, snb, phaser, solve, arp/warp, coot, o, pymol, phenix, hkl2000
|
| Hardware |
4 dual core Opteron cluster running parallelised XDS |
| Backup Devices |
Firewire disk, USBII, DVD, rsync |
|
|